Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints
|Updated:2026-03-12
|
Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints
Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints
解放军医学杂志(英文版)2004年第5期 页码:316-319
Affiliations:
1. Taishan Medical University Tai’an,271000
2. ,China
3. Tai’an Central Hospital Tai’an,271000
Author bio:
Funds:
DOI:
中图分类号:R816.8
纸质出版:2004
Accepted:
Scan QR Code
Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints[J]. 解放军医学杂志(英文版), 2004,(5):316-319.
[1]张兆臣,张有军,冯承强,祝远忠,颜世义,刘玉金.Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints[J].Journal of Medical Colleges of PLA,2004(05):316-319.
Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints[J]. 解放军医学杂志(英文版), 2004,(5):316-319.DOI:
[1]张兆臣,张有军,冯承强,祝远忠,颜世义,刘玉金.Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints[J].Journal of Medical Colleges of PLA,2004(05):316-319.DOI:
Exploration of computerized image processing in underexposed and overexposed X-rays of bones and joints
摘要
Abstract
<正>Objective: To study the effective computerized image processing of underexposed and overexposed X-rays of bones and joints. Methods: Ninety-nine conventional X-ray images (82 were overexposed and 17 were underexposed)
scanned by an UMAX Astra 4000U Scanner
were converted into digital images on the basis of their analog images. A computerized imaging processing program consisting of five functional modules such as Contrast Stretch
Fast Flourier Transform (FFT)
Image Smoothing Modules
Inverse Fast Flourier Transform (IFFT) and Nonlinear Transform performed image contrast stretch and smoothing. Three senior doctors from hospital image sections made their evaluation of all the processed images. Results:Of 82 overexposed films
71 met the clinical requirements after image processing
and 11 were unable to be applied to clinical diagnosis
accounting for 87% and 13% respectively. Of the other 17 underexposed X-ray images
11 met the clinical requirements while 6 were not
making a percentage of 64 and 35.
关键词
Keywords
references
The trial reading is over, you can activate your VIP account to continue reading.