Role of acoustic interface layer during high intensity focused ultrasound therapeutics
|Updated:2026-03-12
|
Role of acoustic interface layer during high intensity focused ultrasound therapeutics
Role of acoustic interface layer during high intensity focused ultrasound therapeutics
解放军医学杂志(英文版)2008年第4期 页码:223-227
Affiliations:
Author bio:
Funds:
the Development Plan for Innovation Teams of Ministry of Education (2005-33);the National Natural Science Foundation of China (30471653);the Natural Science Foundation of Chongqing (2006BA5020)
DOI:
中图分类号:R454
纸质出版:2008
Accepted:
Scan QR Code
Role of acoustic interface layer during high intensity focused ultrasound therapeutics[J]. 解放军医学杂志(英文版), 2008,(4):223-227.
[1].Role of acoustic interface layer during high intensity focused ultrasound therapeutics[J].Journal of Medical Colleges of PLA,2008(04):223-227.
Role of acoustic interface layer during high intensity focused ultrasound therapeutics[J]. 解放军医学杂志(英文版), 2008,(4):223-227.DOI:
[1].Role of acoustic interface layer during high intensity focused ultrasound therapeutics[J].Journal of Medical Colleges of PLA,2008(04):223-227.DOI:
Role of acoustic interface layer during high intensity focused ultrasound therapeutics
摘要
Abstract
After interface layer was simulated by the magnetic nano-particles in the egg white phantom
high intensity focused ultrasound (HIFU) at the same dosage was introduced to radiate the phantom in different depths to blow the acoustic interface layer to mimic "point" exposure. The results showed that the volumes of biological focal region (BFR) were enlarged when the acoustic focal region (AFR) is close with interface layer. This meant that the magnetic nano-particles enhanced the therapeutic efficiency of HIFU. When the distance of the AFR from the interface layer was 10 mm
the size and shape of the BFR were similar with those of the control group
but a larger lesion at the interface
which was harmful for treatment
was observed. When the distance of the AFR to the interface layer increased to 30 mm
the size and shape of the BFR were also similar to those of the control group. When the thickness of the interface layer diminished
the utility of enhancement decreased. Continuous increase of the safe area for treatment and decrease of the utility of enhancement were observed along with the abatement of the thickness of the interface layer